for
Solid State Lab
Thayer School of Engineering
by
Chia-Lun Tsai
1.Start up( Power on)
* Close the main valve.
* Turn on the Main power switch
* Turn on the RF power and pump
Wait at least 10 minutes for the RF power supply to worm up before operate the machine.
* Turn on the gas flow control unit
(make sure valve switches(O2 and CF4) is at the off position
before turn on the main switch)
* Turn on the Vacuum Gauge
* Turn on the cooling pump
2 Pre-set vacuum pressure before loading(optional)
* Make sure Vent valve is close
* Turn on the main valve slowly
* Wait until pressure reach 300 mmTorr
* Open CF4 gas bottle If CF4 gas is needed
* Turn on CF4 valve switch
* Use mechanical valve to adjust the flow
* Adjust the valve until pressure is settle to a desire pressure
* If necessary, turn on the O2 valve switch and adjust as above
3. Loading
* Switch off the CF4 and O2 valve and leave the adjusted mechanical valve.
* Close the main valve
* Vent the chamber
* Load wafer on the electrode and put the cover back on.
* Close the vent valve
* Open main valve slowly to pump down pressure
* Wait until pressure reach below 300 mmTorr
* Switch on O2 and CF4 valve separately
* Wait until chamber is pump down to desire pressure
4. RIE precess
* Turn on the plasma and adjust power level
* Adjust matching network so that reflect power is Min.
* RIE in process
* After process is done, turn off plasma
* Switch off CF4 and O2 valve
* Turn off the main valve
* Vent the chamber slowly
* Take out wafer
* Repeat process begin from Step 3 Loading if needed
5. Before you leave
* Turn off CF4 and O2 mechanical valve
* Turn off CF4 bottle
* Pump down the chamber to less than 1000 mmTorr
* Close the main valve
* Switch off RF power and pump
* There a vent valve behind the main valve, open it
* Close the vent valve after the pipe reach atmosphere
* Turn off cooling pump
* Switch off everything else you see is on.
About Flow Control Unit
Fig.1
Fig.2